Page 24
5.2 Measuring DC milliAmps
5.2.1 Refer to Section 3.0, Preparation For Use, and Section 4.0,
General Operation, before attempting any measurements.
5.2.2 Select the desired DC mA range by rotating the Range
Switch to the desired position. Place the Code Function Switch in
the NORM position. The position of the Signal Select Switch has no
effect in this mode.
5.2.3 Attach the test leads to the Test Set. The red lead is con-
nected to the VOM Jack, and the black lead is connected to the
COM Jack.
5.2.4 Connect the test leads in series with the circuit to be mea-
sured.
5.2.5 Read the current on the appropriate scale as noted in the
table on the following page. If the NEG Indicator lights, then the
input current is reversed from the polarity of leads.
Range Leads Read Multiply
Switch Red_ Black Scale Divide / Direct
15 mADC VOM COM 0 - 15 Direct reading.
60 mADC VOM COM 0 - 60 Direct reading.
300 mADC VOM COM 0 - 300 Direct reading.
Vedere la pagina 23